3. Measuring Loads
Keep the measuring posture vertical and the pressing load constantas much as possible. The basic timing to read is at the time of gettingthe maximum value, but the reading after a certain past time isanother method. The timing of measurement is sometime very important and should be handled very carefully specially with a material on which the reading value starts decreasing just after pressing.
Not with a manual loading measurement, but with a measurementwith some support instrument that can generate a constant load to eliminate the dispersion caused by the difference of operator’s skill, the pressure force to the durometer will vary according to the type of durometer as follows concretely:
- Type A, E, B, O, old A = 10N (approx. 1000gf)
- Type D, C, DO, old C = 50N (approx. 5000gf)
- Type E2 = 5N (approx. 500gf)
- Type 00 = 4N (approx. 400gf)
- Type FO = Self weight measurement
These are the values of force applied to the durometer to obtain a firm contact of the Presser Foot with the specimen according to the spring load each type has.
TECLOCK supplies the CONSTANT LOADER e.g. “TECLOCK GX -610”, a product to supply the durometer with 10 N or 50 N by optional weights.